SENSe Subsystem
SENSe[1]]|SENSe2:CORRection:CFACtor|GAIN[1][:INPut][:MAGNitude]
SENSe[1]]|SENSe2:CORRection:CFACtor|GAIN[1][:INPut][:MAGNitude] <numeric_value>
This command is used to enter a gain correction value for the calibration
factor. The power meter corrects every measurement by this factor to
compensate for the gain.
Either CFACtor and GAIN1 can be used in the command—both have an
identical result. Using GAIN1 complies with the SCPI standard whereas
CFACtor does not but may make your program easier to understand.
Syntax
SENS
1
SENS2
Parameters
Item
numeric_value
(for CFACtor and
GAIN1)
1. For example, a gain of 60% corresponds to a multiplier of 0.6 and a gain of 150%
corresponds to a multiplier of 1.5.
9-22
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
CORR
:GAIN
Description/Default
A numeric value.
• DEF: the default value is 100%.
• MIN: 1%.
• MAX: 150%.
EPM-P Series Power Meters Programming Guide
:INP
1
:MAGN
:CFAC
Space
numeric_value
DEF
MIN
MAX
Space
?
Range of Values
1 to 150 PCT