DISPlay Subsystem
DISPlay[:WINDow[1]|2]:METer:UPPer <numeric_value>
Parameters
Item
numeric_value
Example
DISP:WIND2:MET:UPP 20
Reset Condition
On reset, the upper scale limit is set to 20 dBm.
Query
DISPlay[:WINDow[1]|2]:METer:UPPer? [MIN|MAX]
The query returns the current setting of the analog meter's upper scale
limit or the value associated with MIN and MAX. The format of the response
is <NR3>. The units in which the results are returned is dependent on the
current setting of UNIT:POWer and CALCulate:RELative:STATe as
shown in the previous table.
Query Example
DISP:WIND2:MET:UPP?
5-20
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description/Default
A numeric value for the analog meter
upper scale limit:
• DEF: the default is 20 dBm.
• MIN: -150 dBm.
• MAX: 230 dBm.
Units used are determined by the
current setting of UNIT:POWer and
CALCulate:RELative:STATe as
shown in Table 5-4.
This command sets the lower
window's analog meter upper scale
limit.
This command queries the upper
scale limit set on the analog meter
in the lower window.
EPM-P Series Power Meters Programming Guide
Range of Values
-150 to 230 dBm
DEF
MIN
MAX