Getting The Best Speed Performance - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

Power Meter Remote Operation

Getting the Best Speed Performance

Getting the Best Speed Performance
This section discusses the factors that influence the speed of operation
(number of readings/sec) of an EPM-P Series power meter.
The following factors are those which have the greatest effect upon
measurement speed (in no particular order):
• The selected measurement rate, i.e. NORMal, DOUBle, FAST.
• The sensor being used.
• The trigger mode (for example, free run, trigger with delay etc.).
• The output format: ASCii or REAL.
• The units used for the measurement.
• The command used to take a measurement.
In addition, in FAST mode there are other influences which are described
in "Fast Mode", on page 1-74.
The following paragraphs give a brief description of the above factors and
how they are controlled from SCPI.
Measurement Rate
There are three possible speed settings NORMal, DOUBle and FAST. These
are set using the SENSe:MRATe command and can be applied to each
channel independently (E4417A only).
In NORMal and DOUBle modes, full instrument functionality is available
and these settings can be used with all sensors. FAST mode is available
only for E-Series sensors and averaging, limits and ratio/difference math
functions are disabled.
Refer to "Specifications" in the EPM-P Series Power Meters User's Guide to
see the influence of these speed settings on the accuracy and noise
performance of the power meter.
EPM-P Series Power Meters Programming Guide
1-71
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com