SERVice:BIST:CALibrator <boolean>
This command enables/disables the calibrator self-test during power-up. It
can be used to disable the self-test if it incorrectly indicates failure—for
example a long sensor cable may cause an incorrect failure.
Syntax
SERV
Example
SERV:BIST:CAL OFF
Reset Condition
On reset, the value is set to OFF.
Query
SERVice:BIST:CALibrator?
The query enters a 1 or 0 into the output buffer indicating the status of
the self-test.
• 1 is returned when the self-test is enabled.
• 0 is returned when the self-test is disabled.
Query Example
SERV:BIST:CAL?
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Space
:BIST
:CAL
SERVice Subsystem
SERVice:BIST:CALibrator <boolean>
0|OFF
1|ON
ONCE
?
This command disables the
calibrator self-test during
power-up.
This command queries whether
the self-test is enabled or disabled.
15-3