Trigger[:Sequence]:Level:auto <Boolean> - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

TRIGger Subsystem
TRIGger[:SEQuence]:LEVel:AUTO <boolean>
TRIGger[:SEQuence]:LEVel:AUTO <boolean>
This command enables/disables automatic setting of the trigger level.
When this command is set to:
Note
The command is accepted for TRIGger[:SEQuence[1]] (channel A)
only, for both single and dual channel power meters.
Syntax
TRIG
Example
TRIG:SEQ:LEV:AUTO 0
Reset Condition
On reset the value is set to ON.
Query
TRIGger[:SEQuence]:LEVel:AUTO?
The query enters a 1 or 0 into the output buffer indicating the status of
TRIGger[:SEQuence]:LEVel:AUTO.
13-28
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
ON, automatic setting of the trigger level is enabled.
OFF, automatic setting of the trigger level is disabled.
ONCE, automatic setting of the trigger level is enabled for one
trigger event only. The value is then set to OFF.
:SEQ
:LEV
1 is returned when it is ON.
0 is returned when it is OFF.
EPM-P Series Power Meters Programming Guide
Space
:AUTO
0|OFF
1|ON
ONCE
This command disables the
automatic setting of the trigger
level for channel A.
?