SENSe Subsystem
[SENSe[1]]|SENSe2:TRACe:TIME <numeric_value>
[SENSe[1]]|SENSe2:TRACe:TIME <numeric_value>
This command sets the duration of the trace for E9320 sensors which are
set to NORMal mode. To set the E9320 sensor to NORMal mode, refer to the
command
SENSe[1]]|SENSe2:DETector:FUNCtion <character_data>
on page 9-41.
Syntax
SENS
SENS2
Parameters
Item
numeric_value
Example
SENS2:TRAC:TIME 0.5
Reset Condition
On reset, the duration is set to 100 us.
Query
SENSe[1]]|SENSe2:TRACe:TIME?
The query returns the current duration of the trace.
9-64
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
1
TRAC
:TIME
Description/Default
The duration of the trace in seconds.
• DEF: the default value is 100 us.
Units are resolved to 1 ns.
EPM-P Series Power Meters Programming Guide
Space
numeric_value
DEF
?
Range of Values
10 us to 1 second
DEF
This command sets the duration of
the trace to 0.5 seconds for
channel B.