Measurement Commands
MEASure[1]|2|3|4 Commands
MEASure[1]|2|3|4 Commands
The MEASure? commands configure the power meter to perform a power
measurement with the given measurement function, relative mode
setting, range and resolution then makes the measurement. The format of
the result is set by FORM[:READ][:DATA]. Refer to chapter 6, Format, for
further information.
MEASure? is a compound command which is equivalent to:
2-56
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
For the E4416A the MEASure? query is equivalent to:
ABORt
CONFigure
READ?
For the E4417A carrying out a single channel measurement the
MEASure? queries are equivalent to:
ABORt1
CONFigure
READ1?
or
ABORt2
CONFigure
READ2?
For the E4417A carrying out a difference measurement the
READ:DIFFerence? queries are equivalent to:
ABORt1
ABORt2
CONFigure:DIFFerence
READ:DIFFerence?
For the E4417A carrying out a ratio measurement the
READ:RATio? queries are equivalent to:
ABORt1
ABORt2
CONFigure:RATio
READ:RATio?