SENS:BW:VID:LOW
SENS:SWE1:OFFS:TIME:0.00002
SENS:SWE1:TIME0.00052
INIT:CONT ON
TRIG:SOUR INT
TRIG:LEV:AUTO OFF
TRIG:LEV -20.00DBM
TRIG:DEL 0.00002
TRIG:HOLD 0.004275
TRIG:HYST 3.0
DISP:WIND1:TRACE:LOW -55
DISP:WIND1:TRACE:UPP 20
SENS:TRAC:OFFS:TIME -0.00004
SENS:TRAC:TIME 0.0007
DISP:WIND1:FORM TRACE
DISP:WIND2:FORM DNUM
CALC2:FEED1 "POW:AVER ON SWEEP1" Lower window upper
CALC4:FEED1 "POW:PTAV ON SWEEP1" Lower window lower
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Making Measurements on Wireless Communication Standards
Power Meter Remote Operation
Only send this command if
using an E9323A or E9327A
Sets gate1 start point to
20 μs after the trigger
Sets gate1 length to 520 μs
Puts meter in "wait for
trigger" state
Selects internal trigger
Turn off auto leveling for
trigger
Sets trigger level to
-20.0 dBm
Actual trigger to occur 20 μs
after trig level detected
Sets trigger hold-off to
4.275 ms
Sets Hysteresis to 3 dB
Sets trace display minimum
power to -55 dBm
Sets trace display maximum
power to +20 dBm
Trace starts 40 μs before
trigger point
Trace span set to 700 μs
Assigns upper window to a
trace display
Assigns lower window to a
dual numeric display
display line to show average
power using timing defined
by gate1
display line to show
peak-to-average ratio using
timing defined by gate1