Agilent Technologies EPM-P Series Programming Manual page 156

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

CONFigure[1]|2|3|4 Commands
The CONFigure commands are used on the specified
window/measurement to set:
The CONFigure commands do not make the power measurement after
setting the configuration. Use READ?, or alternatively use INITiate
followed by a FETCh? to make the measurement.
The CONFigure command also applies the following defaults to the
channel(s) which are in the specified window (the channel(s) in the
window are specified in the <source list> parameter):
Default Settings
INITiate:CONTinuous OFF
TRIGger:SOURce IMMediate
TRIGger:DELay:AUTO ON
SENSE:AVERage:COUNt:AUTO ON
SENSE:AVERage:STATe ON
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The expected power level being measured.
The resolution of the window/measurement.
The channel(s) on which the measurement is to be made.
Measurement Commands
CONFigure[1]|2|3|4 Commands
Description
Sets the power meter to make one trigger
cycle when INITiate is sent.
When TRIG:SOUR is set to BUS or HOLD,
sets the power meter to make the
measurement immediately a trigger is
received.
Enables automatic delay before making
the measurement.
Enables automatic filter length selection.
Enables averaging.
2-9

Advertisement

Table of Contents
loading

Table of Contents