MEASure[1]|2|3|4[:SCALar][:POWer:AC]? [<expected_value>[,<resolution>[,<source list>]]]
MEASure[1]|2|3|4[:SCALar][:POWer:AC]? [<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window's measurement function to single
channel with relative mode off, aborts, configures the window then
initiates channel A or B, calculates the measurement result and places the
result on the bus.
Syntax
MEAS
1
:SCAL
2
3
Space
4
Parameters
Refer to "Optional Parameters", on page 2-3 for additional details on the
parameters in this command.
Item
expected_value
(for the expected
power level)
resolution
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:POW
:AC
?
,
expected_value
DEF
Description/Default
A numeric value for the expected
power level. The units of
measurement are dBm and W. The
default units are defined by
UNIT:POWer.
A numeric value for the resolution.
If unspecified the current
resolution setting is used.
Measurement Commands
,
resolution
source list
DEF
Range of Values
sensor dependent
1
DEF
2
1 to 4
1.0, 0.1, 0.01, 0.001
1
DEF
2-57