Measurement Commands
READ[1]|2|3|4[:SCALar][:POWer:AC]:RELative? [<expected_value>[,<resolution>[,<source list>]]]
Error Messages
• INITiate:CONTinuous must be set to OFF, otherwise error -213, "INIT
ignored" occurs.
• If TRIGger:SOURce is set to BUS or HOLD, error -214, "Trigger
deadlock" occurs.
• If the expected value and resolution parameters are not the same as the
current expected value and resolution settings on the specified window,
error -221, "Settings conflict" occurs.