Agilent Technologies EPM-P Series Programming Manual page 404

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

Parameters
Item
numeric_value
1. When a channel is set to 20 or 40 readings/second, TRIG:COUNt is set
automatically to 1.
Example
SPE 40
Reset Condition
On reset, the speed is set to 20 readings/second.
Query
[SENSe[1]]|SENSe2:SPEed?
The query returns the current speed setting, either 20, 40 or 200.
Query Example
SPE?
Error Messages
• If <speed> is not 20, 40 or 200, error -224 "Illegal parameter value"
occurs.
• If an E-Series power sensor is not connected and the
<numeric_value> is set to 200 readings/second, error -241 "Hardware
missing" occurs.
• If TRIG:SOUR is set to EXTernal or INTernal[[1]|2], error -221
"Settings Conflict" occurs.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
[SENSe[1]]|SENSe2:SPEed <numeric_value>
Description/Default
A numeric value for the measurement
speed in readings per second.
The default is 20.
This command sets the channel A
speed to 40 readings/second.
This command queries the current
speed setting for channel A.
SENSe Subsystem
Range of Values
1
20
1
40
200
9-51