Calibration Factor Block Layout

Calibration Factor Block Layout
Calibration Factor Block Layout
The following tables provide information on the calibration factor block
layout for E4410 Series, E9300 Series, E9320 Series and N8480 Series
sensors (excluding Option CFT). The information relates to service
commands described in Chapter 15.
Calibration Factor Block Layout: E4410 Series Sensors
E4410 Series Sensors:
No.
Calibration Factor
Bytes
Block Layout
Header:
Power, low
2
Power, high
2
Number of freq. points
2
Bytes per freq. point
1
Freq. LSB weight
2
Header Total:
Cal Factor Table:
Frequency (point '0')
4
1
Cal factor (low power)
2
1
Cal factor (high power)
2
These table entries are repeated as shown for each frequency point
Frequency (point 'N')
4
1
Cal factor (low power)
2
A-2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Contents
Data Format
Data Range
-
7.8 (signed)
-127.9 to +127.9
-
7.8 (signed)
-127.9 to +127.9
-
16 bit integer
-
-
1000
-
9
-
32 bit fixed
0 to Fhpb* (2^32)
-
2.14
0.25 to 3
-
2.14
0.25 to 3
-
32 bit fixed
0 to Fhpb* (2^32)
-
2.14
0.25 to 3
Units
dBm
Power for low power flatness.
dBm
Power for high power
flatness.
-
None
-
None
Number of bytes in cal factor
value at each frequency and
power level.
Hertz
Fhbp (Freq. Hz per bit).
1 KHz per bit for the cal
factor: 1KHz x 2^32 =
4.3E+12 = 4300 GHz range
None
Fhpb = Freq Hz per bit
None
Power (in watts) is divided by
this value.
None
Power (in watts) is divided by
this value.
None
Fhbp = Freq Hz per bit
None
Power (in watts) is divided by
this value.