Power Meter Remote Operation
Using Sensor Calibration Tables
Each string parameter returned indicates the name, type and size of a
stored sensor calibration table:
• <string>,<type>,<size>
The <string>, <type> and <size> are all character data. The <type>
is always TABL. The <size> is displayed in bytes.
For example, a sample of the response may look like:
560,8020,"Sensor_1,TABL,220","Sensor_2,TABL,340" ....
The power meter is shipped with a set of predefined sensor calibration
tables. The data in these sensor calibration tables is based on statistical
averages for a range of Agilent Technologies power sensors. These power
sensors are:
• DEFAULT
• 8481A
• 8482A
• 8483A
• 8481D
• 8485A
• R8486A
• Q8486A
• R8486D
• 8487A
For further information on naming sensor calibration tables see "Naming
Sensor Calibration Tables", on page 1-47.
Note
Predefined sensor calibration table is not applicable for N8480 Series
power sensors with Option CFT. Therefore you are required to create a
new sensor calibration table for the sensors when a sensor calibration
table is needed.
1. DEFAULT is a sensor calibration table in which the reference calibration
factor and calibration factors are 100%. This sensor calibration table can be
used during the performance testing of the power meter.
2. The 8482B and 8482H power sensors use the same data as the 8482A.