Service:bist:fpath[1]|2:Reference <Numeric_Value> - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

SERVice:BIST:FPATH[1]|2:REFerence <numeric_value>
This command enters an externally measured sensor DC voltage at
0.0 dBm for use as a reference, for the specified channel. It applies to
E9320 Series power sensors only.
Note
Additional hardware is needed to run this command. For further
information refer to the power meter Service Guide.
Syntax
SERV
:BIST
Parameters
Item
numeric_value
Example
SERV:BIST:FPATH2:REF 1.2
Reset Condition
On reset, the last entered value is retained.
Error Messages
• If no power sensor is connected, error -241 "Hardware missing" occurs.
• If a non E9320 Series power sensor is connected, error -241
"Hardware missing" occurs.
• If FPATH2 is sent to a single channel meter, error -113
"Undefined header" occurs.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SERVice:BIST:FPATH[1]|2:REFerence <numeric_value>
:FPATH
:REF
1
2
Description/Default
An externally measured DC voltage
when 0.0 dBm is applied to the sensor.
This command enters an externally
measured sensor voltage of 1.2 v DC
to be used as a reference, for
channel B.
SERVice Subsystem
Space
numeric_value
Range of Values
0.1 to 2.4 v
15-5

Advertisement

Table of Contents
loading

Table of Contents