SENSe Subsystem
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:TIME <numeric_value>
Query Example
SENS2:SWE2:TIME?
Error Messages
• If the command is used when a non E9320 sensor is connected, the
error -241, "Hardware missing" occurs.
• If the command is used when an E9320 sensor is connected and set to
AVERage mode rather than NORMal mode, the error -221, "Settings
Conflict" occurs.
9-56
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
This command queries the length
of the time-gated period for
channel B and gate 2.