Service:sensor[1]|2:Power:average:maximum? - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

SERVice Subsystem

SERVice:SENSor[1]|2:POWer:AVERage:MAXimum?

SERVice:SENSor[1]|2:POWer:AVERage:MAXimum?
This query returns the maximum average power that can be measured by
the currently connected sensor. It is applicable to E-Series sensors only.
Maximum average power information is stored in the sensor's EEPROM.
Syntax
SERV
Example
SERV:SENS:POW:AVER:MAX?
Error Messages
• If no sensor is connected, error -241, "Hardware missing" occurs.
• If a non E-Series is connected, error -241 "Hardware missing" occurs.
• If the E-Series sensor currently connected does not contain the
necessary information in EEPROM, error -241 "Hardware missing"
occurs.
15-20
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:SENS
1
:POW
2
EPM-P Series Power Meters Programming Guide
:AVER
:MAX
This query returns the maximum
average power that can be
measured by the E-Series sensor
currently connected to channel A.
?