Agilent Technologies EPM-P Series Programming Manual page 464

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

SYSTem Subsystem
SYSTem:COMMunicate:SERial:TRANsmit:BAUD <numeric_value>
SYSTem:COMMunicate:SERial:TRANsmit:BAUD <numeric_value>
This command sets the baud rate for both the transmitter and the
receiver. The baud rate for the transmitter and receiver are tied together
and can either be set by this command or the equivalent receiver
command [RECeive]:BAUD.
Syntax
SYST
:COMM
Parameters
Item
numeric_value
Example
SYST:COMM:SER:TRAN:BAUD 38400
Reset Condition
On reset, the baud rate is unaffected.
11-18
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:TRAN
:SER
:BAUD
Description/Default
A numeric value for the baud rate.
the default value is 9600.
DEF:
50.
MIN:
115200.
MAX:
Agilent E4416A/E4417A Programming Guide
Space
numeric_value
DEF
MIN
MAX
Space
?
MIN
MAX
Range of Values
50, 75, 110, 150,
300, 1200, 1800,
2000, 2400, 3600,
4800, 7200, 9600,
19200, 38400,
57600, 115200
DEF
MIN
MAX
This command sets the transmitter
and receiver baud rate to 38400.

Advertisement