Service:sensor[1]|2:Frequency:maximum? - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

SERVice:SENSor[1]|2:FREQuency:MAXimum?

This query returns the maximum frequency that can be measured by the
currently connected sensor. It is applicable to E-Series sensors only.
Maximum frequency information is stored in the sensor's EEPROM.
Syntax
SERV
Example
SERV:SENS2:FREQ:MAX?
Error Messages
• If no sensor is connected, error -241, "Hardware missing" occurs.
• If a non E-Series is connected, error -241 "Hardware missing" occurs.
• If the E-Series sensor currently connected does not contain the
necessary information in EEPROM, error -241 "Hardware missing"
occurs.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SERVice:SENSor[1]|2:FREQuency:MAXimum?
:SENS
1
:FREQ
2
SERVice Subsystem
:MAX
?
This query returns the maximum
frequency that can be measured by
the E-Series sensor currently
connected to channel B.