STATus Subsystem

STATus:OPERation:MEASuring[:SUMMary]

STATus:OPERation:MEASuring[:SUMMary]
The operation status measuring summary register set contains
information on the measuring status of the power meter.
The following bits in these registers are used by the power meter:
Bit
Number
0
1
2
3-15
These bits are set when the power meter is taking a measurement.
These bits are cleared when the measurement is finished.
Syntax
STAT
Note
The operation status measuring summary is available only in average mode.
Refer to the Agilent N1911A/1912A P-Series Power Meters Programming
Guide, Page 791/793, example 2, on measurement polling when using a
VEE Programme.
10-14
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Decimal
Definition
Weight
-
Not used
2
Channel A MEASuring Status
4
Channel B MEASuring Status (E4417A only)
-
Not used
:MEAS
:OPER
EPM-P Series Power Meters Programming Guide