SENSe Subsystem
[SENSe[1]]|SENSe2:MRATe <character_data>
Parameters
Item
character_data
1. When a channel is set to NORMal or DOUBle, TRIG:COUNt is set automatically to 1.
2. FAST mode cannot be used over RS232 or RS422 serial interfaces.
Example
MRAT DOUBle
Reset Condition
On reset, the speed is set to NORMal.
Query
[SENSe[1]]|SENSe2:MRAT?
The query returns the current speed setting, either NORMal, DOUBle or
FAST.
Query Example
MRAT?
Error Messages
• If <character_data> is not set to NORMal, DOUBle or FAST, error -224
"Illegal parameter value" occurs.
• If an E-Series power sensor is not connected and <character_data> is
set to FAST, error -241 "Hardware missing" occurs.
• If TRIG:SOUR is set to EXTernal or INTernal[[1]|2], error -221
"Settings Conflict" occurs.
9-46
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description/Default
A numeric value for the measurement
speed:
• NORMal: 20 readings/second.
• DOUBle: 40 readings/second.
• FAST: up to 1000 readings/second.
The default is NORMal.
This command sets the channel A
speed to 40 readings/second.
This command queries the current
speed setting for channel A.
EPM-P Series Power Meters Programming Guide
Range of Values
1
NORMal
1
DOUBle
2
FAST