CALibration Subsystem
CALibration[1]|2:RCFactor <numeric_value>
CALibration[1]|2:RCFactor <numeric_value>
This command is used with 8480 Series power sensors and N8480 Series
power sensors with Option CFT to set the reference calibration factor of
the specified channel. Reference calibration factors can also be set using
sensor calibration tables. The power meter uses the most recently set
reference calibration factor.
Syntax
CAL
Parameters
Item
numeric_value
Example
CAL1:RCF 98
Reset Condition
On reset, the reference calibration factor is set to 100%.
4-12
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Space
:RCF
1
2
?
Description/Default
A numeric value:
DEF:
the default is 100%.
MIN
: 1%.
MAX:
150%.
EPM-P Series Power Meters Programming Guide
numeric_value
DEF
MIN
MAX
Space
MIN
MAX
Range of Values
1.0 to 150.0 PCT
DEF
MIN
MAX
This command enters a reference
calibration factor of 98% to
channel A.