Bit Definitions - Standard Event Register - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

Table 1-4: Bit Definitions - Standard Event Register
Bit
Number
0
1
2
3
4
5
6
7
The standard event register is cleared when you:
• send a *CLS (clear status) command.
• query the event register using the *ESR? (event status register)
command.
The standard event enable register is cleared when you:
• cycle the instrument power.
• execute a *ESE 0 command.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Decimal
Value
1
Operation Complete
All overlapped commands following an
command have been completed.
2
Not Used. (Always set to 0.)
4
Query Error
A query error occurred, refer to error numbers 410 to
440 in the User's Guide.
8
Device Error
A device error occurred, refer to error numbers 310 to
350 in the User's Guide.
16
Execution Error
An execution error occurred, refer to error numbers
211 to 241 in the User's Guide.
32
Command Error
A command syntax error occurred, refer to error
numbers 101 to 161 in the User's Guide.
64
User request.
128
Power On
Power has been turned off and on since the last time
the event register was read or cleared.
Power Meter Remote Operation
Status Reporting
Definition
*OPC
1-87