Agilent Technologies EPM-P Series Programming Manual page 184

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

FETCh[1]|2|3|4[:SCALar][:POWer:AC]:RATio:RELative? [<expected_value>[,<resolution>[,<source
Item
1. The mnemonic
parameter used in the command sub-systems. The parameters must be entered in
the specified order. If parameters are omitted, they will default from the right. The
parameter DEFault is used as a place holder. Specifying
value unchanged.
significant digits. When the measurement result is logarithmic 1 to 4 represents a
resolution of 1, 0.1, 0.01 and 0.001 respectively.
Example
FETC:RAT:REL?
Error Messages
• If the last measurement on either channel is not valid error -230, "Data
corrupt or stale" occurs. A measurement is valid after it has been
initiated. It becomes invalid when either a reset occurs or any
measurement parameter, for example frequency, is changed.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description/Default
If it is unspecified the current
resolution setting is used. If a
value is entered it should
correspond to the current
resolution setting otherwise an
error occurs.
channels used to calculate the
If unspecified and the current
window setup is a ratio
is used, otherwise it defaults to
channel A/B (E4417A) or A/A
(E4416A).
DEF
means DEFault. This is not equivalent to the DEFault
This command queries the relative ratio
measurement on the upper window.
Measurement Commands
list>]]]
Range of Values
2
1.0, 0.1, 0.01, 0.001
1
DEF
3
(@1),(@2)
3
(@2),(@1)
(@1),(@1)
3
(@2),(@2)
leaves the parameter
DEF
2-37