MEMory Subsystem
MEMory:STATe:DEFine <character_data>,<numeric_value>
MEMory:STATe:DEFine <character_data>,<numeric_value>
This command is used to associate a name with a save/recall register
number.
Syntax
:STAT
MEM
Parameters
Item
character_data
numeric_value
Example
MEM:STAT:DEF "SETUP1",4
Query
MEMory:STATe:DEFine? <string>
The query returns the register number for the given register name.
Query Example
MEM:STAT:DEF? "SETUP1"
7-20
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Space
character_data
:DEF
?
Description/Default
Details the register name. A maximum of
12 characters can be used.
A numeric value (<NRf>) for the register
number.
EPM-P Series Power Meters Programming Guide
,
numeric_value
Space
character_data
Range of Values
A to Z (uppercase)
a to z (lowercase)
0 - 9
_ (underscore)
0 to 9
This command names register 4
SETUP1.
This command queries the register
number of SETUP1.