TRIGger Subsystem
TRIGger[:SEQuence]:HYSTeresis <numeric_value>
TRIGger[:SEQuence]:HYSTeresis <numeric_value>
This command sets:
Note
The command is accepted for TRIGger[:SEQuence[1]] (channel A)
only, for both single and dual channel power meters.
Syntax
TRIG
Example
TRIG:SEQ:HYST 0.1
Parameters
Item
numeric_value
Reset Condition
On reset the value is set to 0 dB.
13-24
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
How far a signal must fall below TRIG:LEVel before a rising edge
can be detected.
How far a signal must rise above TRIG:LEVel before a falling
edge can be detected.
:SEQ
:HYST
Description/Default
How far a signal must fall/rise before
a rising or falling edge can be
detected.
the default value is 0 dB.
DEF:
Units are resolved to 0.05 dB.
EPM-P Series Power Meters Programming Guide
Space
numeric_value
DEF
?
This command sets the value to
2 dB for channel A.
Range of Values
0 to 3 dB
DEF