Parameters
Item
string
Example
CALC3:FEED2 "POW:AVER ON
SWEEP2"
Reset Condition
On reset, data_handle is set to :POW:AVER.
Query
CALCulate[1]|2|3|4:FEED[1]2?
The query returns the current value of the string.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description
The input measurement type to be fed
to the specific input on the CALC
block:
• PEAK: peak power.
• PTAV: peak to average.
• AVER: average.
Values may be followed by
ON SWEEP[1]|2|3|4
where the numeric specifies the gate
to be used for the feed. For example:
"POW:PEAK ON SWEEP2"
If ON SWEEP[1]|2|3|4 is not
supplied, the gate used is left
unchanged.
A feed of "" (empty string) disables
the CALC block and switches off that
display line.
This command selects the input for
FEED2 of CALC block CALC3 to be
average power, using gate 2. The
channel from which the feed is taken
is determined by CALC:MATH:EXPR.
CALCulate Subsystem
CALCulate[1]|2|3|4:FEED[1]|2 <string>
Range of Values
"POW:PEAK"
"POW:PTAV"
"POW:AVER"