*RCL <NRf>
The *RCL <NRf> (ReCaLl) command restores the state of the power meter
from the specified save/recall register. An instrument setup must have
been stored previously in the specified register.
Syntax
*RCL
Parameters
Type
NRf
Error Message
• If the register does not contain a saved state, error -224, "Illegal
parameter value" occurs.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Space
NRf
Description/Default
The number of the register to be recalled.
IEEE488.2 Command Reference
*RCL <NRf>
Range of Values
1 - 10