FETCh[1]|2|3|4[:SCALar][:POWer:AC]? [<expected_value>[,<resolution>[,<source list>]]]
Item
source list
1. The mnemonic
parameter used in the command sub-systems. The parameters must be entered in
the specified order. If parameters are omitted, they will default from the right. The
parameter DEFault is used as a place holder. Specifying
value unchanged.
2. When the measurement result is linear this parameter represents the number of
significant digits. When the measurement result is logarithmic 1 to 4 represents a
resolution of 1, 0.1, 0.01 and 0.001 respectively.
Example
FETC2:POW:AC?
Error Messages
• If the last measurement is not valid error -230, "Data corrupt or stale"
occurs. A measurement is valid after it has been initiated. It becomes
invalid when either a reset occurs or any measurement parameter, for
example frequency, is changed.
• If the expected_value and resolution parameters are not the same as
the current expected value and resolution setting on the specified
window, error -221, "Settings conflict" occurs.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description/Default
The channel which the command
is implemented on. If unspecified
the current window setup is used.
However, on the E4417A, if the
window shows a ratio or difference
measurement, the upper window
defaults to channel A and the
lower window to channel B.
DEF
means DEFault. This is not equivalent to the DEFault
This command queries the lower
window's measurement result.
Measurement Commands

Range of Values

(@1)
(@2) (E4417A only)
DEF
leaves the parameter
2-25