SYSTem Subsystem
SYSTem:PRESet <character_data>
BLUetooth
The Bluetooth set-up provides:
The measurement is started by detecting the Bluetooth RF burst using
the internal RF level trigger. The internal trigger is set to –20dBm.
Time-gating is used to measure the peak and average power in a single
Bluetooth DHI data burst which lasts for 366us. The DHI burst does not
occupy a full Bluetooth timeslot, which lasts for 625us.
The following table shows the power meter presets when
<character_data> is set to BLUetooth. Commands not listed are
preset according to their DEFault values (for further information refer to
Table 11-1):
11-44
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Average power in a Bluetooth DH1 data burst.
Peak power in the same burst.
Display of RF pulse in one timeslot.
Agilent E4416A/E4417A Programming Guide