The Status Byte Summary Register
The status byte summary register reports conditions from other status
registers. Query data waiting in the power meter's output buffer is
immediately reported through the "message available" bit (bit 4). Clearing
an event register clears the corresponding bits in the status byte summary
register. Reading all messages in the output buffer, including any pending
queries, clears the message available bit.
Table 1-3: Bit Definitions - Status Byte Register
Bit
Number
0
1
2
3
4
5
6
7
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Decimal
Weight
1
Not Used (Always set to 0)
2
Device Status Register summary bit.
One or more bits are set in the Device Status
Register (bits must be "enabled" in enable register)
4
Error/Event Queue
8
Questionable Status Register summary bit.
One or more bits are set in the Questionable Status
Register (bits must be "enabled" in enable register).
16
Data Available
Data is available in the power meter's output buffer.
32
Standard Event
One or more bits are set in the Standard Event
register (bits must be "enabled" in enable register).
64
Request Service
The power meter is requesting service (serial poll).
128
Operation Status Register summary bit.
One or more bits are set in the Operation Status
Register (bits must be "enabled" in enable register).
Power Meter Remote Operation
Status Reporting
Definition
1-85