CALibration Subsystem

CALibration Subsystem
CALibration Subsystem
The CALibration command subsystem is used to zero and calibrate the
power meter. It is also used to set the reference calibration factor for the
power sensor which is being used.
The numeric suffix of the CALibration command refers to a specific
channel:
Zeroing and calibration of the power meter is recommended:
The following CALibration commands are overlapped commands:
An overlapped command allows the instrument to continue parsing and
executing subsequent commands while it is still executing.
Keyword
CALibration[1]|2
[:ALL]
[:ALL]?
:AUTO
:ECONtrol
:RCALibration
:RCFactor
:ZERO
4-2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
CALibration1 represents channel A.
CALibration2 represent channel B.
This command does not apply to the single channel E4416A power
meter and results in the error "Header suffix out of range."
When a 5°C change in temperature occurs.
When you change the power sensor.
Every 24 hours.
Prior to measuring low level signals. For example, 10 dB above the
lowest specified power for your sensor.
CAL:ALL
CAL:AUTO
CAL:ZERO:AUTO
Parameter Form
<boolean>|ONCE
:STATe
<boolean>
<boolean>
<numeric_value>
:AUTO
<boolean>|ONCE
:NORMal
:AUTO
<boolean>|ONCE
EPM-P Series Power Meters Programming Guide
Notes
Page
[event; no query]
page 4-3
[event;query]
page 4-5
page 4-7
page 4-9
page 4-10
[non-SCPI]
page 4-12
page 4-14
page 4-15