System:help:headers? - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

SYSTem:HELP:HEADers?

This query returns a list of all SCPI commands supported by the
instrument.
Data is returned in IEEE 488.2 arbitrary block program data format as
follows:
#xyyy..yddd................ddd<LF>
Example: if there are 12435 data bytes, y = 12435 and x = 5
Each point in the trace is represented as an IEEE 754 32 bit floating point
number, made up of four bytes in the data block. The MS byte is
transmitted first. Each complete block is terminated by a line feed.
Commands are listed in alphabetical order.
Syntax
SYST
Example
SYST:HELP:HEAD?
Agilent E4416A/E4417A Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The number of data bytes (d)
contained in the block.
The number of y digits
Signifies the start of the block
:HELP
:HEAD
SYSTem Subsystem
SYSTem:HELP:HEADers?
Line feed character
signifies the end of the block
Data bytes
?
This command returns the SCPI
commands supported by the
instrument.
11-27