DISPlay Subsystem
DISPlay[:WINDow[1]|2]:METer:LOWer <numeric_value>
Parameters
Item
numeric_value
Example
DISP:WIND2:MET:LOW 10
Reset Condition
On reset, the lower scale limit is set to -70 dBm.
Query
DISPlay[:WINDow[1]|2]:METer:LOWer? [MIN|MAX]
The query returns the current setting of the analog meter's lower scale
limit or the value associated with MIN and MAX. The format of the response
is <NR3>. The units in which the results are returned is dependent on the
current setting of UNIT:POWer and CALCulate:RELative:STATe as
shown in the previous table.
Query Example
DISP:MET:LOW?
5-18
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description/Default
A numeric value for the analog meter
lower scale limit.
• DEF: the default value is -70 dBm.
• MIN: -150 dBm.
• MAX: 230 dBm.
The default units are defined by
UNIT:POWer and
CALCulate:RELative:STATe.
This command sets the lower
window's analog meter lower scale
limit.
This command queries the lower
scale limit set on the analog meter
in the upper window.
EPM-P Series Power Meters Programming Guide
Range of Values
-150 to 230 dBm
DEF
MIN
MAX