SYSTem Subsystem
SYSTem:COMMunicate:SERial:CONTrol:RTS <boolean>
SYSTem:COMMunicate:SERial:CONTrol:RTS <boolean>
This command is used to set the serial interface (RS232 or RS422) RTS
signal line either high always (ON) or low always (OFF). In addition, the
RTS signal line can be tied to the condition of the interface receiver buffer
(IBFull) resulting in the line going high when the receiver is ready to
accept data and low when the receiver buffer is full (not ready for data).
Also, with RTS in IBFull mode, the transmitter will be inhibited when
CTS is low.
Syntax
SYST
Example
SYST:COMM:SER:CONT:RTS ON
Reset Condition
On reset, the condition of the RTS signal line is unaffected.
Query
SYSTem:COMMunicate:SERial:CONTrol:RTS?
The query returns the current RTS signal line setting, ON, OFF or IBFull.
Query Example
SYST:COMM:SER:CONT:RTS?
11-8
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:COMM
:SER
:CONT
Agilent E4416A/E4417A Programming Guide
Space
:RTS
ON
OFF
IBFull
?
This command sets the RTS signal
line high (always).
This command queries the setting
of the RTS signal line.