CALCulate[1]|2|3|4:LIMit:STATe <boolean>
This command enables/disables the test limits for the specified window.
Syntax
CALC
Example
CALC2:LIM:STAT 1
Reset Condition
On reset, limit checking is disabled.
Query
CALCulate[1]|2|3|4:LIMit:STATe?
The query enters 1 or 0 into the output buffer indicating the status of the
limits testing feature for the specified window/measurement.
Query Example
CALC1:LIM:STAT?
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:LIM
:STAT
1
2
3
4
1 is returned when limits testing is enabled.
0 is returned when limits testing is disabled.
CALCulate Subsystem
CALCulate[1]|2|3|4:LIMit:STATe <boolean>
Space
0|OFF
1|ON
?
This command enables the limit
checking function for the lower
window upper measurement.
This command queries whether
the limit checking function for the
upper window/upper
measurement is on or off.