SYSTem:COMMunicate:SERial[:RECeive]:BAUD <numeric_value>
This command sets the baud rate for both the receiver and the
transmitter. The baud rate for the receiver and transmitter are tied
together and can either be set by this command or the equivalent
transmitter command :TRANsmit:BAUD.
Syntax
SYST
:COMM
Parameters
Item
numeric_value
Example
SYST:COMM:SER:REC:BAUD 38400
Reset Condition
On reset, the baud rate is unaffected.
Agilent E4416A/E4417A Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SYSTem:COMMunicate:SERial[:RECeive]:BAUD <numeric_value>
:REC
:SER
:BAUD
Description/Default
A numeric value for the baud rate.
the default value is 9600.
DEF:
50.
MIN:
115200.
MAX:
SYSTem Subsystem
Space
numeric_value
DEF
MIN
MAX
Space
?
MIN
MAX
Range of Values
50, 75, 110, 150,
300, 1200, 1800,
2000, 2400, 3600,
4800, 7200, 9600,
19200, 38400,
57600, 115200
DEF
MIN
MAX
This command sets the receiver
and transmitter baud rate to
38400.