SERVice Subsystem

SERVice:SENSor[1]|2:FREQuency:MINimum?

SERVice:SENSor[1]|2:FREQuency:MINimum?
This query returns the minimum frequency that can be measured by the
currently connected sensor. It is applicable to E-Series sensors only.
Minimum frequency information is stored in the sensor's EEPROM.
Syntax
SERV
Example
SERV:SENS1:FREQ:MIN?
Error Messages
• If no sensor is connected, error -241, "Hardware missing" occurs.
• If a non E-Series is connected, error -241 "Hardware missing" occurs.
• If the E-Series sensor currently connected does not contain the
necessary information in EEPROM, error -241 "Hardware missing"
occurs.
15-18
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:FREQ
:SENS
1
2
EPM-P Series Power Meters Programming Guide
:MIN
?
This query returns the minimum
frequency that can be measured by
the E-Series sensor currently
connected to channel A.