Output:ttl[1]|2:Feed <String> - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

OUTput Subsystem
OUTPut:TTL[1]|2:FEED <string>
OUTPut:TTL[1]|2:FEED <string>
This command controls which limit test drives a given TTL output. There
are two TTL outputs:
Both can be connected to any of the CALC subsytems (that is the upper
window upper/lower measurement or the lower window upper/lower
measurement) using OUTput:TTL:FEED <string>.
Syntax
Parameters
TTL
[1|2]
1
2
8-14
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Output 1 (TTL1).
Output 2 (TTL2).
OUTP
:TTL
1
2
String
"CALC1|2|3|4:LIM:LOW"
"CALC1|2|3|4:LIM:UPP"
"CALC1|2|3|4:LIM:LOW,
1
CALC1|2|3|4:LIM:UPP"
"CALC1|2|3|4:LIM:LOW"
"CALC1|2|3|4:LIM:UPP"
"CALC1|2|3|4:LIM:LOW,
1
CALC1|2|3|4:LIM:UPP"
Space
:FEED
?
Description
TTL Output 1 asserted indicates a
lower limit fail on the appropriate
window/measurement.
TTL Output 1 asserted indicates an
upper limit fail on the appropriate
window/measurement.
TTL Output 1 asserted indicates that
either the lower or upper limit failed on
the appropriate window/measurement.
TTL Output 2 asserted indicates a
lower limit fail on the appropriate
window/measurement.
TTL Output 2 asserted indicates an
upper limit fail on the appropriate
window/measurement.
TTL Output 2 asserted indicates that
either the lower or upper limit failed on
the appropriate window/measurement.
EPM-P Series Power Meters Programming Guide