SENSe Subsystem
[SENSe[1]]|SENSe2:SPEed <numeric_value>
[SENSe[1]]|SENSe2:SPEed <numeric_value>
This command sets the measurement speed on the selected channel. The
speeds available are 20, 40 and 200 readings/second.
Note
This command is included for compatibility purposes only. It has the
same purpose as [SENSe[1]]|SENSe2:MRATe
<NORMal|DOUBle|FAST> (with 20 mapping to NORMal, 40 to DOUBle
and 200 to FAST), which should be used in preference.
When a channel is set to 200 readings/second the following couplings
occur:
[SENSe[1]]|SENSe2:AVERage:STATe
[SENSe[1]]|SENSe2:CORRection:DCYCle:STATe
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe
CALCulate[1|2|3|4]:GAIN:STATe
CALCulate[1|2|3|4]:RELative:STATe
CALCulate1|3:MATH:EXPRession
CALCulate2|4:MATH:EXPRession
1. This change only occurs on the channel specified in the SENSe:SPEed
command. When the specified channel is changed from 200 readings/second to
either 20 or 40 readings/second the settings that were in place when 200
readings/second mode was entered are restored.
2. This change occurs when either channel is set to 200 readings/second. When
both channels are changed from 200 readings/second to either 20 or 40
readings/second the settings that were in place when 200 readings/second mode
was entered are restored.
3. Applicable to the E4417A dual channel power meter only.
Syntax
SENS
SENS2
9-50
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Command
:
Space
SPE
1
EPM-P Series Power Meters Programming Guide
Status
1
OFF
1
OFF
1
OFF
2
OFF
2
OFF
"(SENSe1)"
3
"(SENSe2)"
numeric_value
?

Advertisement

Table of Contents
loading

Table of Contents