Advertisement

SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut][:MAGNitude] <numeric_value>
Reset Condition
On reset, GAIN2 is set to 0.00 dB.
Query
[SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut][:MAGNitude]? [MIN|MAX]
The query returns the current setting of the channel offset or the values
associated with MIN and MAX.
Query Example
CORR:GAIN2?
Error Messages
• If a loss/gain correction value is entered using
[SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut][:MAGNitude]
while [SENSe[1]]|SENSe2:SPEed is set to 200, the error -221,
"Settings Conflict" occurs. However, the correction value is set but the
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe command is not
automatically set ON.
• The SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut]
[:MAGNitude] command can be used for the 8480 Series power sensors
and N8480 Series power sensors with Option CFT when no sensor
calibration table has been set up. If a sensor calibration table is selected
the error -221, "Settings Conflict" occurs.
9-40
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
This command queries the current
setting of the channel offset on
channel A.
EPM-P Series Power Meters Programming Guide