To use sensor calibration tables you:
1. Edit a sensor calibration table if necessary.
2. Select the sensor calibration table.
3. Enable the sensor calibration table.
4. Zero and calibrate the power meter. The reference calibration
factor used during the calibration is automatically set by the
power meter from the sensor calibration table.
5. Specify the frequency of the signal you want to measure. The
calibration factor is automatically set by the power meter from the
sensor calibration table.
6. Make the measurement.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Power Meter Remote Operation
Using Sensor Calibration Tables
1-43