CALibration[1]|2:ZERO:NORMal:AUTO <boolean>
This command provides a quick way of zeroing the NORMAL path of an
E9320 Series sensor. The average only path is unaffected. The command
cannot be used to zero a non-E9320 Series sensor.
The command causes the power meter to perform its zeroing routine on
the specified channel when ONCE is selected. This adjusts the power meter
for a zero power reading with no power supplied to the power sensor.
The 0|OFF parameter is only required for the query response and is
ignored in the command. If 1|ON is selected, it causes the error -224,
"Illegal parameter value" to occur.
The command assumes that the power sensor is not connected to a power
source.
Syntax
CAL
Example
CAL2:ZERO:NORM:AUTO ONCE
Reset Condition
On reset, automatic zeroing is disabled.
Query
CALibration[1]|2:ZERO:NORMal:AUTO?
The query always returns a value of 0.
Error Messages
• If zeroing was not carried out successfully the error -231, "Data
Questionable; ZERO ERROR" occurs. If you are using a dual channel
power meter, the error message specifies which channel the zeroing
failed on.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
CALibration[1]|2:ZERO:NORMal:AUTO <boolean>
:ZERO
:NORM
1
2
CALibration Subsystem
Space
:AUTO
This command causes the power meter to
perform a zeroing routine on channel B.
0|OFF
ONCE
?
4-15