[SENSe[1]]|SENSe2:V2P ATYPe|DTYPe

This command is used to select the type of linearity correction that will be
applied to the channel sensors being used. For most 8480 Series power
sensors, the correct (A type or D type) linearity correction table is
automatically selected. However, for the V8486A and W8486A sensors the
automatic selection must be overridden and the D type (diode) correction
selected.
syntax
Example
SENS2:V2P DTYP
Reset Condition
On reset, the linearity correction is set for A type.
Query
[SENSe[1]]|SENSe2:V2P?
The query returns the current type of linearity correction being used
(A type or D type).
Query Example
SENS:V2P?
Error Messages
If no sensor is connected or the sensor is not an A type, the error -241,
"Hardware missing" occurs.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:
SENS
1
SENS2
[SENSe[1]]|SENSe2:V2P ATYPe|DTYPe
Space
V2P
ATYP
DTYP
?
This command selects the D type
linearity correction to be applied to
channel B.
This command queries which
linearity correction type is
currently being used on channel A.
SENSe Subsystem
9-67