Agilent Technologies EPM-P Series Programming Manual page 237

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

CALCulate Subsystem
CALCulate[1]|2|3|4:LIMit:UPPer[:DATA] <numeric_value>
Parameters
Item
numeric_value
Example
CALC2:LIM:UPP:DATA 5
Reset Condition
On reset, both channels are set to +90.00 dBm or +90 dB.
Query
CALCulate[1]|2|3|4:LIMit:UPPer[:DATA]? [MIN|MAX]
The query returns the current setting of the upper limit or the values
associated with MIN and MAX for the specified window/measurement.
Query Example
CALC2:LIM:UPP:DATA?
3-22
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description/Default
A numeric value for the upper
test limit:
• DEF: the default is
+90.00 dBm or +90 dB.
• MIN: -150 dBm or -180 db.
• MAX: +230 dBm or +200 dB.
This command enters an upper
limit for the lower window/upper
measurement depending on the
window's units as follows:
dBm = 5 dBm
W = 5 W
dB = 5 dB
% = 5%
This command queries the setting
of the upper limit for the lower
window/upper measurement.
EPM-P Series Power Meters Programming Guide
Range of Values
-150 to +230 dBm or
-180 to +200 dB
DEF
MIN
MAX