SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2[:SELect] <string>
Query
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:[SELect]?
The name of the selected table is returned as a quoted string. If no table is
selected an empty string is returned.
Query Example
CORR:CSET1?
Error Messages
• If <string> is not valid, error -224, "Illegal parameter value" occurs.
• If a table called <string> does not exist, error -256, "File name not
found" occurs.
• When a sensor calibration table is selected, the power meter verifies
that the number of calibration points defined is one more than the
number of frequency points defined. When a frequency dependent offset
table is selected, the power meter verifies that the number of offset
points defined is equal to the number of frequency points defined. If this
is not the case, error -226, "Lists not the same length" occurs.
• If the CSET1 command is used when an E-Series power sensor or N8480
Series power sensor (excluding Option CFT) is connected the error -241,
"Hardware missing" occurs.