MEMory Subsystem
MEMory:TABLe:GAIN[:MAGNitude] <numeric_value>{,<numeric_value>}
Parameters
Item
numeric_value
Example
MEM:TABL:SEL "Sensor_1"
MEM:TABL:GAIN 97,99.5,97.4
Query
MEMory:TABLe:GAIN[:MAGNitude]?
The query returns a list of calibration factor/offset points for the currently
selected table.
Query Example
MEM:TABL:GAIN?
Error Messages
If more than 81 calibration factors for sensor calibration tables, or 80
offsets for frequency dependent offset tables are in the list, error -108,
"Parameter not allowed" occurs.
If a table is not specified using the MEMory:TABLe:SELect command,
the data cannot be entered and error -221, "Settings conflict" occurs.
If any of the calibration/offset factors are outside of the allowed range,
error -222, "Data out of range" occurs.
7-28
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Description/Default
A numeric value for the calibration/
offset factors. The units are PCT.
This command enters a reference
calibration factor of 97% and
calibration factors of 99.5% and
97.4% into the sensor calibration
table.
This command queries the
calibration factor/offset in the
current table.
EPM-P Series Power Meters Programming Guide
Range of Values
1.0 to 150.0

Advertisement