Agilent Technologies EPM-P Series Programming Manual page 50

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

SENS:FREQ:900MHZ
SENS:BW:VID:HIGH
SENS:BW:VID:LOW
SENS:SWE1:OFFS:TIME:0.00002
SENS:SWE1:TIME0.00052
INIT:CONT ON
TRIG:SOUR INT
TRIG:LEV:AUTO OFF
TRIG:LEV -20.00DBM
TRIG:DEL 0.00002
TRIG:HOLD 0.004275
DISP:WIND1:TRACE:LOW -35
DISP:WIND1:TRACE:UPP 20
SENS:TRAC:OFFS:TIME -0.00004
SENS:TRAC:TIME 0.0007
DISP:WIND1:FORM TRACE
DISP:WIND2:FORM SNUM
CALC2:FEED1 "POW:AVER ON SWEEP1" Lower window to show
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Making Measurements on Wireless Communication Standards
Power Meter Remote Operation
Sets the measurement
frequency to 900 MHz
Only send this command if
using an E9321A or E9325A
Only send this command if
using an E9323A or E9327A
Sets gate1 start point to
20 μs after the trigger
Sets gate1 length to 520 μs
Puts meter in "wait for
trigger" state
Selects internal trigger
Turn off auto leveling for
trigger
Sets trigger level to
-20.0 dBm
Actual trigger to occur 20 μs
after trig level detected
Sets trigger hold-off to
4.275 ms
Sets trace display minimum
power to -35 dBm
Sets trace display maximum
power to +20 dBm
Trace starts 40 μs before
trigger point
Trace span set to 700 μs
Assigns upper window to a
trace display
Assigns lower window to a
single numeric display
average power using timing
defined by gate1
1-25

Advertisement

Table of Contents
loading

Table of Contents