Bit
Number
4
5
10
11
12
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Meaning
EVENts Causing Bit Changes
MEASuring
This is a summary bit for the Operation MEASuring
Register.
• SET: When the power meter is taking a
measurement.
• CLEARED: When the measurement is finished.
Waiting for
This is a summary bit for the Operation TRIGger
TRIGger
Register.
• SET: When the power meter enters the "wait for
trigger" state.
• CLEARED: When the power meter enters the
"idle" state.
SENSe
This is a summary bit for the Operation SENSe
Register.
• SET: When the power meter is reading data from
the E-Series power sensor EEPROM.
• CLEARED: When the power meter is not reading
data from the E-Series power sensor or N8480
Series power sensor EEPROM.
Lower Limit
This is a summary bit for the Lower Limit Fail
Fail
Register.
• SET: If a measurement is made and either a
channel or window lower limit test fails.
• CLEARED: If a measurement is made and the
lower limit test is not enabled or the test is
enabled and passes.
Upper Limit
This is a summary bit for the Upper Limit Fail
Fail
Register.
• SET: If a measurement is made and either a
channel or window upper limit test fails.
• CLEARED: If a measurement is made and the
upper limit test is not enabled or the test is
enabled and passes.
Power Meter Remote Operation
Status Reporting
1-91