Bluetooth: Power Meter Presets - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

Table 11-11: BLUetooth: Power Meter Presets
Command
Frequency
[SENS[1]]|SENS2:FREQ[:CW|:FIX]
Sensor measurement mode
[SENS[1]]|SENS2:DET:FUNC
Sensor video bandwidth setup
[SENS[1]]|SENS2:BAND|BWID:VID
Gate setup
[SENS[1]]|SENS2:SWE[1]|2|3|4
:OFF:TIME
[SENS[1]]|SENS2:SWE[1]|2|3|4
:TIME
Trigger setup
TRIGger[:SEQuence[1]|2]:SOUR
INIT:CONT
TRIG[:SEQ]:LEV:AUTO
TRIG[:SEQ]:LEV
TRIG[:SEQ]:SLOP
TRIG[:SEQ]:DEL
TRIG[:SEQ]:HOLD
Step detection
[SENSe[1]]|SENS2:AVER:SDET
Trace setup
DISP[:WIND[1]|2]:TRAC:LOW
DISP[:WIND[1]|2]:TRAC:UPP
[SENS[1]]|SENS2:TRAC:OFF
:TIME <numeric_value>
[SENS[1]]|SENS2:TRAC:TIME
<numeric_value>
Agilent E4416A/E4417A Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SYSTem:PRESet <character_data>
Setting
+2400.000 MHz
Frequency setting
NORM
Measurement mode
E9321A/25A: DEF
Sensor video bandwidth
E9322A/26A: HIGH
E9323A/27A: MEDium
Gate 1: 0.2 us
Delay between trigger point
Gates 2 - 4: 0
and time gated period.
Gate 1: 366 us
Length of time gated period
Gates 2 - 4: 0
for time gated measurements.
INT1
Trigger source set up and
acquisition mode continuous
ON
triggering
OFF
Disable automatic setting of
the trigger level
-20 dBm
Power level
POS
Trigger event recognized on
the rising edge of a signal
0
Delay between recognition of
trigger event and start of a
measurement
650 us
Trigger holdoff
1
Step detection enabled
+20 dBm
Maximum power
-35 dBm
Minimum power
-50 us
Delay between delayed
trigger point and the start of
the trace
3.8 ms
Length of the trace
SYSTem Subsystem
Comments
11-45

Advertisement

Table of Contents
loading

Table of Contents